Entrance A and Entrance B areas 500mV, 5V, 50V, 500V, 750V accuracy ±(0.5% + 5) Common Mode Re
Entrance A and Entrance B
areas
500mV, 5V, 50V, 500V, 750V
accuracy
±(0.5% + 5)
Common Mode Rejection (CMRR)
>100 dB at DC, >60 dB at 50, 60 or 400 Hz
Display scope
5,000 counting steps
areas
500mV, 5V, 50V, 500V, 750V
Accuracy for 5% to 100% of range (DC coupled)
DC to 60 Hz (V AC+DC)
±(1% + 10)
±(1% + 10)
Accuracy for 5% to 100% of range (AC or DC coupled)
60Hz to 20kHz
±(2.5% + 15)
DC suppression (VAC only)
>50dB
Common Mode Rejection (CMRR)
>100 dB at DC
Display scope
5000 counting steps, measured value is independent of the crest factor.
Display types
Max Peak, Min Peak, or Peak to Peak
areas
500mV, 5V, 50V, 500V, 2200V
accuracy
Max Peak or Min Peak
5% of the range end value
10% of the range end value
Display scope
500 counting steps
areas
123B: 1Hz, 10Hz, 100Hz, 1kHz, 10kHz, 100kHz, 1MHz, 10MHz, and 50MHz
Frequency range
15 Hz (1 Hz) to 50 MHz with continuous automatic adjustment
Accuracy at 1 Hz to 1 MHz
±(0.5% + 2)
Display scope
Display size up to 10,000
Maximum value
50,000 rpm
accuracy
±(0.5% + 2)
Area
2% to 98%
Frequency range
15 Hz (1 Hz) to 30 MHz with continuous automatic adjustment
Frequency range
15 Hz (1 Hz) to 30 MHz with continuous automatic adjustment
Display scope
1000 counting steps
with current clamp
areas
such as V DC, V AC, V AC+DC or peak value
0.1mV/A, 1mV/A, 10mV/A, 100mV/A, 400mV/A, 1V/A, 10mV/mA
as with V DC, V AC, V AC+DC, or PEAK (the inaccuracy of the current clamp must be added)
Area
200°C/div (200°F/div)
Scaling factor
1mV/°C and 1mV/°F
accuracy
as with V DC (the inaccuracy of the temperature sensor must be added)
0dBV
1V
0dBm (600Ω /50Ω)
1 mW related to 600 Ω or 50 Ω
dB for
V DC, V AC or V AC+DC
Display scope
1000 counting steps
Area
1 to 10
Display scope
90 counting steps
Types of measurement
A to B, B to A
Area
0 to 359 degrees
resolution
1 degree
Configurations
1-phase or 3-phase with symmetrical load (3-phase: only fundamental component, only AUTOSET operation)
Power factor (PF)
Ratio between active power and apparent power: 0.00 to 1.00
Active power (watts)
RMS measurements of corresponding samples from input A (voltage) and input B (current)
999 counting steps
Apparent power (VA)
V eff x A eff
999 counting steps
Reactive power (VAR)
√((VA)²-W²)
999 counting steps
Purpose
For measuring pulse width modulated signals, e.g. B. at the outputs of frequency converters
principle
Measurements show the effective voltage based on the average of sample points over a range of periods of the fundamental frequency
accuracy
as with Veff for sine signals